Youjoy Best-22L-B: undocumented error stops sterilization cycle

What happened

A steam sterilizer displayed fault A01, which is not listed in the available documentation — the manual only describes E-prefixed codes (E01 there means a temperature-sensor fault, temperature above 250 °C as reported in the case).

What was checked

Cause

Poor feed-water quality: contamination had built up in the reservoir, and the unit signalled it as A01.

What was done

The engineers drained the water, cleaned the reservoir and refilled it — the fault cleared. Regular reservoir cleaning was recommended (for example with a wet/dry vacuum suitable for wet pick-up), as reported in the case.

Safety

Before opening the reservoir, isolate the sterilizer from the mains, let it cool down and release any residual pressure; handle the drained water as potentially contaminated, with gloves and eye protection.

Lesson: When a fault code is missing from the manual, do not force the closest documented code to fit — verify the machine's actual condition such as water quality and contamination, since the cause is often maintenance-related rather than a sensor fault.