Ziemer · фемтолазер Ziemer (Femto LDV, точно не указано)

Ziemer Femto LDV: touchscreen failure, then no boot

What happened

In this reported case, a Ziemer femtosecond laser (Femto LDV family) first lost touchscreen response, and later the unit stopped booting altogether.

What was checked

Cause

The engineers found that the real fault in this case was in the power supply (a cheap multifunctional module was also mentioned), not in the motherboard initially suspected.

What was done

Reported in the case: independent engineers carried out component-level repair of the power supply as an alternative to an expensive board replacement; by that time the facility had already written the unit off, so the correct diagnosis came too late to return the laser to service.

Safety

Work inside a laser power supply only with the unit isolated from mains and with capacitors discharged and verified dead — stored energy there can be lethal; after any such repair, complete electrical safety testing and a supervised laser output check before any return to clinical use.

Lesson: Measure the power supply rails and its modules thoroughly before ordering an expensive motherboard — a failing supply can mimic controller and display faults, and a wrong board order can cost a device its life.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.