How to document the competency of biomedical engineers for JCI and ISO audits
2026-11-21
1. What auditors look for
When a hospital or a service company is audited, the question about the people who maintain medical equipment is simple: can you show that each of them is qualified for the equipment they work on? Auditors look for evidence of training, evidence that competence was assessed, and the dates.
The important point is who owns this evidence. It is the organisation, not the course provider and not the manufacturer. A certificate from any course is one document in the file; the organisation decides that a person is competent for a given type of equipment and keeps the record that says so.
2. Why it is hard for independent and hospital teams
Manufacturer training is the reference for a specific system, but it is expensive, has long waiting lists and covers one vendor. A hospital team that maintains monitors, ventilators, infusion pumps and imaging from many brands cannot send everyone to every factory course.
Most real learning happens on the job: a senior engineer shows a junior how to run a test or find a fault. That training is valuable, but it usually leaves no record, so at the audit there is nothing to show for it.
3. A simple record that holds up
For every engineer and every type of equipment, keep four things: the training taken and its hours; how competence was assessed (an exam, an observed task or both); the date and the person who signed off; and the date a refresher is due.
| Engineer | Equipment | Training and hours | Assessment | Signed off (date, by) | Refresher due |
|---|---|---|---|---|---|
| Engineer A | Patient monitors | Patient Monitor Service, Level 1, online course with certificate | Final exam; observed preventive maintenance | date, senior engineer | date |
| Engineer B | Ventilators | Manufacturer course for the model in use | Observed performance test | date, head of department | date |
What ISO 13485 and ISO 9001 expect from training records
ISO 13485 applies to organisations involved in the life cycle of medical devices, including those that service them. Hospitals are usually not certified to it, but independent service companies often are. Its human resources requirement (clause 6.2) asks the organisation to:
- determine the competence needed for people whose work affects product quality;
- provide training or other action to reach that competence;
- evaluate whether the training was effective;
- make sure people know how their work matters;
- keep records of education, training, skills and experience.
ISO 9001 has a similar requirement on competence (clause 7.2), including keeping documented evidence of it.
In practice, an auditor will ask for the record behind each name: what competence the role needs, what training was done, how its effectiveness was checked, and when. The table in this guide covers all of these points. The part teams most often miss is the effectiveness check: write down how you confirmed that the training worked, for example an exam score plus an observed task.
Always check the current text of the standard your organisation is certified to. This section is a summary, not a quotation.
Download the template
Training record template (CSV): one row per engineer and equipment type, with columns for training, hours, assessment, effectiveness check, sign-off and next review.
4. Where outside courses help
A structured course with a final exam and a certificate that can be checked online fills the first two columns for the theory part: what was studied, for how many hours, and an independent test of knowledge. It is one input to the record, not the whole record.
The observed practical sign-off stays with the hospital or the service company: a senior engineer watches the person do the work on the real equipment and signs. Keep both in the same file and the record answers the auditor's question.
5. How we do it
Service Engineer Academy gives a company one table of its engineers' courses, exam scores and certificates, with training records exported as PDF or CSV.