Vivid
GE HealthCare Vivid is a family of diagnostic ultrasound systems, mainly used for cardiovascular (echo) imaging and also for general imaging, ranging from portable units to full-size cart-based consoles.
34 field cases from service engineers' discussions (anonymised); solved: 9, probable: 25under engineering review
Faults and how to find them
System does not power on, or does not start from the power button (power LED or standby indicator may be lit; the ultrasound part may stay dark while the PC side is alive)field case
- Faulty PC power supply, e.g. the 3.3 V rail reading too high (seen in the logs on a Vivid E9)
- Faulty main power supply (MPS) that does not start the ultrasound part of the system
- Dried-out capacitors in the power supply, so the 5 V enable signal is missing after the unit has been fully off for a long time (Vivid 7)
- 3.3 V missing on the computer board (Vivid S5)
- Faulty component on the motherboard in the power-on circuit (Vivid E9)
- Dirty contacts on the BEP / control panel side of the power button circuit (Vivid 7 Dimension)
- Read the system logs and the service menu voltage readings (3.3 V, 5 V)
- Check the LEDs on the ultrasound boards, above the motherboard and on the BEP/VER block to see which section has no power
- Measure the voltages at the power supply connector and check whether the relay that feeds the modules switches on
- Trace the power-button signal through the I/O board to the motherboard, drawing the circuit if needed
- Look at the ultrasound board nearest the power supply for burnt capacitors
- Swap in a known-good power supply to confirm or rule it out
- Clean the contacts on the BEP
Solved in one case by replacing the PC power supply that had the 3.3 V overvoltage error; the errors went away. In other cases the likely fixes were replacing the MPS (the system started with a substitute supply), recapping the power supply when the fault appears after long idle periods, cleaning the BEP contacts, or repairing or replacing the motherboard (in one case reheating a bridge chip brought the unit back only temporarily, with a motherboard replacement still pending). Forcing the supply on is a diagnostic step only, not a repair.
Poor image quality or poor Doppler, image artifacts, dark image, noise with no echo, or a static image that does not respond to the probe, on one probe, one port or all probesfield case
- Interference coming in on the mains power cable (the same probes worked fine on another system)
- Fault in a port or the probe board (image came back after swapping probes between ports)
- Faulty Relay board when only the linear probe is affected and convex/sector probes are normal (Vivid 7)
- No high voltage (HV) from the power supply, or a beamformer board fault, when there is no live echo on any probe (Vivid T8)
- Faulty crystal module inside the probe (3Sc artifacts in all modes, probe serial/info lost)
- Unstable TGC slider readings causing a cyclic band on all probes (Vivid E95)
- Move the suspect probes to a known-good system to tell probe faults from system faults
- Swap probes between ports
- Try other rooms and outlets, or run the system from an online UPS
- Check whether all probes are affected or only one probe type
- Reset the software, run the DC offset calibration and the system tests that are available
- Move the TGC sliders to their end positions and use the front-panel diagnostic utility to see whether the slider values stay steady or jump around
- Check the HV supply and the beamformer board when there is no live image
Solved in one case by fitting ferrite rings on the mains cable. In the other cases the likely fixes were repairing or replacing the Relay board (and checking the related resistors), replacing the probe's crystal module, checking the power supply HV and the beamformer board, or investigating the port/probe board further after a port swap brought the image back.
Boot loop, the system will not boot normally (safe modes), or it does not respond to any key except powerfield case
- Probe plugged into the rear tee connector upside down (cable pointing up instead of down) (Vivid iq)
- Mechanically damaged probe connectors: hit against the couch while the system was moved with a probe attached, or worn out over time (Vivid S5/S6)
- Dead backup battery causing boot loops after service tests (Vivid E90)
- Check how every probe is plugged in, including the tee connector on the cart
- Disconnect the probe connector board (PSB) and see whether the system then boots normally
- Take the probe connectors apart and inspect them
- Disconnect the battery
- Reseat the ultrasound and PC boards; reinstall the software only after hardware causes have been ruled out
Solved by plugging the probe in the right way round, and in another case by repairing or replacing the damaged probe connectors. For the boot loop after service tests, disconnecting the battery is the likely fix (a known cause is a dead battery).
Buttons or encoders on the control panel activate by themselves, or characters are typed at randomfield case
- Faulty IC (possibly a solder defect) on the encoder board of the keyboard (Vivid S6)
- Liquid spilled on the keyboard (suspected, not confirmed)
- Inspect the boards for dirt and liquid
- Reseat the ribbon cables
- Run the system without the control panel, then with the encoder board disconnected, to find the source
- Press every key in turn and watch for stuck keys
Solved by replacing the faulty IC on the keyboard board. In the suspected spill case, disconnecting the keyboard and working the keys restored normal operation.
Trackball problems: cursor sticks or jerks along one axis after some time in use, or the trackball well is too dirty to clean from outsidefield case
- Failed diode in the optical trackball (suspected; Vivid S5)
- Dirt in the trackball well
- Find out whether the trackball is optical or mechanical
- Try blowing it out with compressed air
- On the Vivid E9 the trackball is reached from the inside: remove the lower monitor, then the top panel
Cleaning access through the lower monitor and top panel worked on the Vivid E9. For the optical trackball that fails on one axis, replacing the trackball is the likely fix.
Overheating error, or a 'clean the filter' message that stays after cleaningfield case
- Dust in the lower fan unit even though the rest of the system was cleaned and the fans spin (Vivid E9)
- Clogged filters and dust throughout the system (Vivid S70)
- Look at the errors in the service menu
- Check that all fans are running
- Take out the lower fan unit and inspect it
Likely fixed by taking out and cleaning the lower fan unit, or by fully stripping the system, cleaning every filter and blowing it out.
Probes not detected at power-up, built-in test stops at 'DC verification offset', 'incorrect software configuration' message, power supply shows no information or keeps losing its identification data (Vivid S6)field case
- Faulty DC/DC block or power supply whose onboard memory has lost its identification data
- Run the built-in test
- Check what the system shows about the power supply
Likely fixed by replacing the faulty memory chip or the DC/DC block. If the data keeps getting lost, replace the block or update the system software to the latest version, which was said to depend less on this data. Use only manufacturer-approved methods.
Intermittent 'HW id 0' error that clears after a reboot; now and then the screen and keyboard switch off and the system goes to standby (Vivid E9)field case
- Motherboard fault (suggested from a similar case)
- Inspect the cable between the front-end and back-end and its pins
The motherboard is the likely cause. Not confirmed (single case).
Software or operating system problems: system will not start or run properly, Windows is corrupted, the system will not boot from USB for a reinstall, image upside down on the main monitor, or exam settings reset to defaults for every new patientfield case
- Errors in the database partition on the old hard disk (Vivid 7/S5/S6)
- Failed hard disk (Vivid E9)
- Newer PC module with a SATA hard disk that does not see a USB stick or DVD as boot media (Vivid q)
- Screen-rotation hotkeys in the graphics card utility triggered by accident (Vivid S60)
- No user preset saved (Vivid E9)
- Check the hard disk and its database partition
- Look for the original software media: in an envelope under the system's covers (under the black cover or on the left by the CD drive)
- For USB boot problems: use a USB hub with its own power supply, check whether the PC has an IDE or SATA disk, and consider a dual-layer DVD if the image is small enough
- Check the graphics utility hotkeys for screen rotation
- Create a user preset with the required settings
Solved in one case by temporarily changing the computer name in the settings, rebooting and changing it back, after moving to a new disk. The other fixes are likely: reinstall from the original media, replace the hard disk, set the screen rotation back with the hotkeys (a horizontal offset remained), or save a custom preset.
Field notes
Before replacing boards, move the probes to a known-good system and swap ports: this quickly tells probe, port and system faults apart.
Check how probes are physically connected and what state their connectors are in. A probe plugged into the tee upside down, or a cracked connector, can stop the whole system from booting.
Don't move the system with probes plugged in: connectors damaged against the couch were a confirmed cause.
Mains interference can look like a degraded image or Doppler. Try other outlets, ferrite rings or an online UPS before suspecting the front-end.
For no-power faults, work section by section: LEDs, voltages in the service menu and logs, then swap the power supply. Recurring faults after long power-off suggest ageing capacitors.
Dust builds up in the lower fan unit even when the fans spin and the rest of the system has been vacuumed.
After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.