DC series
The Mindray DC series is a family of cart-based diagnostic ultrasound systems built around an embedded PC workstation, used for general imaging and cardiac exams.
24 field cases from service engineers' discussions (anonymised); solved: 8, probable: 16under engineering review
Faults and how to find them
The system will not start. It gives only a short beep and nothing more, tries to power up and immediately shuts down, or starts only after several attempts.field case
- Discharged BIOS (CMOS) battery on the motherboard
- If the battery is fine: a faulty board (I/O board or the main/CPU board)
- Measure the BIOS battery voltage. In one case it read normal (3.20 V) and the fault turned out to be elsewhere.
- Check the battery size before ordering. CR2032 was used in these cases; CR2025 was asked about for one model.
- Measure the power supply output voltages at its output connectors
- Check the main power supply by fitting it to a known-good system
- Work through the troubleshooting section of the service manual
On a DC-3 with only a short beep, replacing the CR2032 BIOS battery fixed it (this needs the housing opened). On other units, replacing the battery and then setting the date and time was recommended as the likely fix. Where the battery tested good and the power supply was confirmed working, a faulty board was suspected, but no confirmed fix was reported.
The system starts unreliably, freezes or glitches, the control panel or trackball does not respond, or USB ports stop working. Sometimes the built-in screen shows 'no signal' while an external monitor works, and a battery warning may appear.field case
- Failed I/O (interface) board. Field reports call this a common failure on DC-3 and DC-N6/T6.
- A failing DC-DC converter on the I/O interface board that cannot carry the load (monitor, control panel and disc drive share one supply)
- Short circuit in the control panel, or failed suppressors on the USB ports
- On one DC-40 (model uncertain), a motherboard fault: the keyboard answered only to probe selection
- Check that the control panel gets power and that its backlight works
- Plug a standard PC USB keyboard into the system to see whether input works at all
- Confirm that the probes still work
- Check the control panel for a short circuit and inspect the USB-port suppressors
- Disconnect the disc drive's power and/or the battery, then restart, to see whether lowering the load changes the behaviour
Probable: replace or repair the I/O interface board (including its DC-DC section). On a DC-N6, disconnecting the disc drive got the system to start once, but the fault came back on the next start. For the DC-40 case with an unresponsive keyboard, replacing the motherboard was recommended. None of these repairs was confirmed in the field.
The system does not power on: it does not boot and the keyboard does not light. On a DC-N6, the 4th indicator lights first and then goes out.field case
- Faulty DC-DC board, which sits above the AC-DC unit
- Check the indicators on the DC-DC board above the AC-DC unit. Three indicators should be lit; if they are not, suspect the DC-DC board.
Probable: a fault on the DC-DC board. No confirmed repair was reported.
Boot problems: a HAL.dll error at Windows start-up, failing to start some of the time, endless loading of Windows or the application, or hanging on the logo screen for a very long time (15 minutes to an hour) before it may finally boot.field case
- A failing hard disk (logs showed the system restarting several times a day before the disk died completely)
- File system errors on the hard disk
- Corrupted application software
- The disk filling up because the temp folder has grown too large
- Review the Windows logs for repeated restarts
- Check free disk space and the size of the temp folder
- On Linux-based systems (Resona / DC-X6), open the boot console (Alt+D, or Ctrl+Alt+F1/F2/F3) to see where the boot stops. If it asks for a file system check, answer 'y'.
- Allow plenty of time for a slow boot before deciding the system is dead
On a DC-7, replacing the hard disk and reinstalling the OS and application fixed it. Access to the disk on the DC-7 needs the whole housing removed. On a DC-30, reinstalling the application software fixed an endless boot, and clearing a bloated temp folder was also recommended. On a DC-X6 with a very slow boot, the likely fix was a file system check followed by moving to or replacing the hard disk.
The system freezes during a particular exam mode (transcranial Doppler on a DC-70 Pro) and must be switched off and on again, or an application recovery install will not complete.field case
- Corrupted or wrong application software version
- The installer version does not match the system or OS (in one case the recovery would not install under Windows 10)
- Confirm the software version and the OS the system runs
- Try a different, correct installer version and do not keep retrying the same failing one
For the transcranial Doppler freeze, reinstalling the software was recommended and is a likely fix, not a confirmed one. For the failed recovery install, installing a suitable version of the software restored normal operation. Get software only from the manufacturer or another authorised source.
Printer problems: the thermal printer keeps printing stored images (about 50 queued, media used up), or the system sees the printer (a Mitsubishi P95D in one case) but nothing prints.field case
- A stuck or overfilled print queue
- A damaged or incorrect printer driver on the supplied disc
- Open the print queue and clear it
- Check the print settings and how the print key is assigned
- Reinstall the printer driver from a freshly downloaded copy rather than from the original disc
- Consider the printer connection type. Connecting by BNC instead of USB was suggested.
Clearing the print queue stopped the endless printing. Downloading the driver again to a USB stick and installing it from there got printing working. Both fixes were confirmed.
Grainy or noisy image: either with iBeam and iClear switched off (DC-40), or in cardiac exams with the SP5-1E phased-array probe (DC-70 Pro).field case
- Normal behaviour: the image gets coarser without the iBeam/iClear processing
- Normal image character of the SP5-1E probe on this system, which looks less 'polished' than on a DC-8
- Compare the image with iBeam/iClear switched on and off
- Compare against another probe (for example a convex probe) on a phantom
- Where possible, compare with a similar probe on another system
In both cases the grain was confirmed to be normal behaviour, not a fault. No repair was needed. Explain the image-processing settings to the user.
Interference (bright bands or gaps) on the image (DC-8)field case
- Electrical interference from the UPS or other nearby equipment
- Run the system self-test
- Unplug the UPS from the mains and check the image again
- Move the system to another room, away from possible interference sources
Probable: the UPS or the surroundings were the source of the interference. Checking the UPS or moving the system was recommended; no outcome was confirmed. In the same case, the SP5-1E probe was reported as not supported on that DC-8 unit (built in 2015).
Damaged video output after the HDMI connector was pulled out and shorted (DC-70)field case
- Damage to the video output stage on the board that carries the HDMI output
- Possible damage to the video scaler
- Inspect the HDMI connector and the board it sits on
- Check whether the scaler was also damaged
Probable: replacing the board with a new one was recommended as the fastest and most reliable option. Repairing the board at component level was mentioned as an alternative. No outcome was reported.
Field notes
For no-start or 'starts only after several tries' faults, check the cheap things first: measure the BIOS battery and the power supply outputs before condemning a board.
The I/O interface board and its DC-DC section are a common weak point. Symptoms include freezes, failure to start, a dead control panel and USB failures, especially on DC-3 and DC-N6/T6.
Repeated unexplained restarts in the Windows logs, or very slow boots, often come before hard disk failure. On DC-7, replacing the disk means removing the whole housing.
Before treating a grainy image as a fault, compare with iBeam/iClear on and off and against another probe on a phantom. On several units this turned out to be normal image character.
Printer faults are usually software: clear the print queue first, and if the original driver disc does not work, reinstall from a freshly downloaded driver.
After any hardware or software repair, run the manufacturer's checkout/verification procedure before the system goes back into clinical use.
After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.