Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Library / CT / SOMATOM
Siemens Healthineers

SOMATOM

Siemens Healthineers SOMATOM is a family of multislice computed tomography (CT) scanners used for routine and contrast-enhanced diagnostic imaging, including angiographic studies.

29 field cases from service engineers' discussions (anonymised); solved: 8, probable: 21under engineering review

02

Faults and how to find them

Images do not reach PACS. Either automatic sending does not happen at all, or only one image of a DICOM series is sent instead of the whole series. The scanner still acquires and reconstructs normally, and the PACS answers ping.field case
Likely causes
  • Sending to a new PACS node is not configured in the protocol editor
  • Network/DICOM adapter settings were lost, for example after moving a backup to different host hardware
  • Corrupted local databases (partial series sent)
Checks, in order
  1. Check the send configuration window and the auto-send rules in the protocol editor for the PACS node
  2. Re-create the DICOM rule and run verification
  3. Verify the connection from both sides, including on the PACS side
  4. Check the network adapter settings, especially after a hardware migration or backup restore
Fix

Field fixes: set up sending for the new PACS in the protocol editor; restore or reconfigure the network adapter settings. In the case where only one image per series was sent, all databases were removed. Before that, the required series were extracted as ISO images through the file section into /service/extraction/.

The operator console software does not load. The system hangs or reports an error partway through boot, syngo stops at the splash screen with the counting numbers and drops to the Windows desktop, or syngo will not start after a hardware change.field case
Likely causes
  • The previous session was not shut down properly. On the next start the software sees the improper-shutdown flag and repairs the damaged program file
  • The SRS/host profile or network settings were disrupted after the DVD drive was replaced and its drive letter changed
  • A problem with the csageneric process (the exact cause was not found)
Checks, in order
  1. Do a full system restart
  2. Reseat the LAN cables at the operator desk, the switch and the scan room. In one case this was not the cause
  3. After a drive-letter change, check the Satellite/local profile and the IP address and hostname in local devices
  4. Check the hard disks
  5. Locate a valid backup or system image before reinstalling
Fix

A full reboot fixed the hang after an improper shutdown, and a reboot is advised if it happens again. After the drive-letter change, switching the Satellite profile to local and entering the IP address and hostname got syngo running again. For syngo dropping to the desktop, the likely path is a disk check, then restoring the system image or reinstalling the software. This was not confirmed.

The host or console PC goes straight into BIOS setup at every power-on. After leaving BIOS it boots and works normally. A similar problem follows BIOS battery replacement.field case
Likely causes
  • A stuck Del key on an extra PS/2 keyboard someone has connected
  • BIOS settings lost or reset after the BIOS battery was replaced or went flat
Checks, in order
  1. Look for extra keyboards or peripherals connected to the host and disconnect them
  2. Before any BIOS work, photograph all current BIOS settings
  3. Measure the BIOS battery (one low battery read 2.9 V) and replace it if needed
  4. Load BIOS defaults, then restore the original settings by hand from the photos
Fix

Confirmed in one case: removing the extra keyboard with the stuck Del key fixed it. After battery replacement, the likely fix is to load defaults and restore the recorded BIOS configuration by hand.

The gantry does not start or respond. On Definition AS: the gantry and console do not switch on, the SCR or 'SR Closed' LED on UMAS/UMAS2 is off, and earlier there were spontaneous gantry/UMAS2 shutdowns. On Emotion: the fans run but buttons do nothing, gantry communication tests fail, the table clicks but does not move, and the gantry display shows '8's.field case
Likely causes
  • The UMAS unit has failed
  • The Stop Report loop signal is lost somewhere in the chain
  • No 24 V supply to UMAS
  • A stuck button on the control panel (Emotion)
  • Suspected in one case but not confirmed: board D301 or the communication board in the ICS (Emotion)
Checks, in order
  1. Check UMAS power and the 24 V indication
  2. Trace the Stop loop with the manual's loop diagram to find where the signal drops
  3. Check the inverter cabinet and how the MCI board behaves
  4. Check the control panel buttons for one that is stuck
  5. Firmware reloading was considered pointless when the UMAS does not boot
Fix

The likely fix for a dead UMAS is to replace the unit. As a temporary measure, disassembling it, renewing the thermal pads and reseating the boards may revive it briefly. On Emotion, a similar case was solved by finding and fixing a stuck control-panel button. This is likely, not confirmed for the case in question.

Signals degrade on detector channels: the signal graph shows worsening on certain modules, and modules fail in a cluster over several months.field case
Likely causes
  • Detector modules degrading with age
  • Temperature swings during long shutdowns, for example air conditioning switched off over holidays
  • Contrast contamination of the detector area
Checks, in order
  1. Compare the event log against the dates when modules failed
  2. Use the per-module signal graph to find the failing modules
  3. Open and visually inspect the detector block, checking for contamination by contrast
  4. Inspect the collimator
Fix

The likely fix is to replace the failing detector modules, choosing them from the signal graph. Keep room climate stable during long shutdowns.

X-ray tube and cooling problems: air in the tube cooling circuit (gurgling when the pump runs, flowing sound when the gantry rotates, intermittent cooling pressure and temperature errors, no visible leaks); short tube life with arcing well before the expected scan-seconds (DURA202); after a tube repair, focusAli and zfocuscheck fail in Tune-up.field case
Likely causes
  • Tube arcing turning oil into gas
  • A leak in the cooling circuit (oil can evaporate without leaving traces)
  • Not enough tube cooling
  • Checkup with getter activation not done regularly
  • Collimator out of adjustment after tube work
Checks, in order
  1. Check the logs for cooling pressure and temperature errors and for arcing
  2. Check that the tube cooling works properly
  3. Check that the checkup procedure (getter activation) is run regularly
  4. After tube work, check and adjust the collimator before repeating Tune-up
Fix

For air in the circuit, the recommendation was to replace the tube and cooling system, under warranty where it applies. For short tube life, ensure proper cooling and run checkup with getter activation regularly; no software limit on tube life was found. For Tune-up failures after a tube repair, collimator adjustment was recommended. All of these are probable fixes.

Patient table problems: the scanner switches off and reboots when the table is raised or lowered; the lift squeaks; horizontal calibrations are lost after power-off; a PMM (Physiological Measurement Module) communication error appears with the table stuck in the lowest position.field case
Likely causes
  • Faulty encoder on the vertical lift motor
  • Brake relay (one participant guessed position ~13)
  • Lift screw not lubricated
  • Flat encoder battery (GM400.E16 encoder)
  • Lost communication with the PMM
Checks, in order
  1. Review the logs and check the vertical lift motor encoder through service access
  2. Check whether horizontal calibrations reset after power-off, and check the encoder battery
  3. Find the lift's lubrication nipple
  4. To reach parts under a fully lowered table, use the manual lift mechanism at the bottom of the table
Fix

Probable fixes: replace the vertical lift motor encoder (a similar case was solved this way), or check the brake relay. Lubricate the lift through its built-in nipple with a non-graphite oil. Replace the encoder battery and check the result.

After the databases are cleaned, the disk-usage indicator (green circle at the bottom of the screen) still shows about 70%.field case
Likely causes
  • The exchangedatabase (intermediate database on drive D) is cluttered
  • A disk problem, for example after a disk replacement
Checks, in order
  1. Check the disks and hardware
  2. Confirm the databases are re-created and images are deleted from all three disks
  3. Find out whether disks were replaced before the problem started
Fix

The likely fix is to stop the cleanb processes in Task Manager, then remove the exchangedatabase on drive D with the utilities as administrator. The alldbremove batch file was also used to clean the databases.

The scanner boots fully, but the console mouse and keyboard do nothing for about 40 minutes, then start working. The mouse is lit and the logs show no errors. Plugged in at the gantry, the peripherals work.field case
Likely causes
  • A faulty USB extender, the USB 3.0 to fiber optic cable
Checks, in order
  1. Review the error logs
  2. Test the peripherals plugged in at the gantry
  3. Check the cable between the ICS and the monitor, and the monitor's USB hub
Fix

The likely fix is to replace the USB extender (USB 3.0 to fiber optic cable), as in a colleague's similar case.

05

Field notes

Before BIOS work, photograph every setting so you can restore it by hand after loading defaults.

Before deleting any database, extract the needed patient series (for example as ISO images to /service/extraction/).

If a fault starts after someone else worked on the system (extra keyboard, new UPS, hardware migration, new PACS), check that change first.

On an MMWP workstation, after reinstalling the software on drive C, run the script that creates the RAID and sets up the disks and directories. Otherwise configuration cannot be finished.

Long holiday shutdowns without climate control were linked to detector module failures. Keep the room environment stable.

For contrast leg studies with no dedicated protocol, the angio runoff protocol was recommended, with bolus monitoring below the aortic bifurcation and a delay set so the scan does not outrun the contrast.

After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.