Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Library / CT / Aquilion
Canon Medical (Toshiba)

Aquilion

The Aquilion is a family of multi-slice CT scanners from Canon Medical (formerly Toshiba). Hospitals use it for routine diagnostic CT, contrast-enhanced CT and CT angiography.

39 field cases from service engineers' discussions (anonymised); solved: 11, probable: 28under engineering review

02

Faults and how to find them

An 'over kV' error ('Over kV (Fixed)') appears during exposure. In one case it showed only at high settings (120 kV / 500 mA, large focus), and lower currents gave high voltage without errors.field case
Likely causes
  • Faulty 24 V power supply that feeds the HV multiplier
  • Dried-out capacitors in the 24 V power supply mounted on the multiplier
Checks, in order
  1. Recalibrate the filament currents first, to rule out a calibration problem
  2. Check the 24 V supply on the HV multiplier under load, not just at idle, and use the generator graphs
  3. Inspect the HV cable, the silicone ring and the tube receptacle
Fix

Solved in two cases. One was fixed by replacing the 24 V power supply that feeds the HV multiplier. The other was fixed by replacing the dried-out capacitors in that supply, then verifying it under load. A spare supply is worth keeping. After the repair, run the manufacturer's generator and image checkout.

An XC error appears after the consoles power up and blocks scanning. After a reset, the tube temperature reads 100%, then 'cools down' and scanning is allowed. A related case showed 'xc error' / 'xc local' with 'filament timeout' and 'starter communication error' in the generator logs.field case
Likely causes
  • Flat battery in the original timekeeper (measured 0.3 V)
  • Fault in the starter (anode rotation drive). This is a probable cause only.
Checks, in order
  1. Measure the timekeeper battery voltage
  2. Decode the generator register using the troubleshooting section of the service manual
  3. Connect to the XC and test it
  4. Check the starter that spins up the anode
Fix

Timekeeper case, solved: replaced the timekeeper, then recalibrated the filament currents. Starter case, probable: inspecting the starter (anode rotation drive) was recommended, and the cause was not confirmed.

An 'over current' error appears right after boot, before warm-up or any exposure, and the logs also show 'over ma'. In another case, a generator error appears from time to time but the system keeps working after the error window is closed.field case
Likely causes
  • HV cable or X-ray tube problem (probable)
  • Faulty temperature sensor on the tube (seen by one engineer)
  • Generator power supply problem (suspected only)
Checks, in order
  1. Check the HV cable at the generator end
  2. Review the generator graphs and match the error to the description in the service manual
  3. Test systematically rather than guessing between the generator and the tube
Fix

Probable only. Start by checking the HV cable at the generator end; the fault is likely in the tube or the HV cable. For the intermittent error, one engineer found a faulty tube temperature sensor, but this was not confirmed for the case discussed. Only qualified staff should work on HV parts, and the manufacturer's checkout applies afterwards.

Patient table drive faults. On one system the servo driver shows error 21 (lost encoder communication) from time to time, the table jams, and switching to manual mode from the gantry panel does not work. On another, the AC servo faults when the table moves backward at high speed, but slow movement works both ways. On a third, the table moves fine by hand but does not move when a scan starts, and the system faults.field case
Likely causes
  • Slipping electromagnetic brake between the motor and the gearbox, so the motor encoder and the gearbox-output encoder disagree (not confirmed)
  • Stiff encoder bearing
  • Inter-turn short in the motor windings, causing an overcurrent fault (mentioned as a similar case)
  • Worn horizontal drive motor assembly
Checks, in order
  1. Remove, clean and refit the encoder, and check that it turns freely by hand
  2. Remove mechanical obstructions; lubricate and adjust the rollers and bearings
  3. Check the infrared sensors
  4. Check the servo driver power supply and reset the KGSTM board
  5. Connect a cable to the AC servo and read its logs, or read the error code from the LED blink pattern
Fix

All three cases are probable. In the stiff-bearing case, greasing the encoder bearing while warming it with a heat gun was followed by no repeat of the error. For the encoder-mismatch case, colleagues recommended replacing the whole drive (servo motor with gearbox). For a table that will not move during scans, replacing the horizontal drive motor assembly was recommended. After any table repair, check the table motion and positioning with the manufacturer's procedure.

Image artifacts. In one case the scout views were fine, but about halfway through a scan the image filled with interference. Other cases showed ring-shaped bright artifacts that persisted after water and air calibration, and grainy noise after NRA and Air calibrations.field case
Likely causes
  • Damaged fibre-optic data cable (for the interference that starts mid-scan)
  • Degraded detectors, or degraded boards/DAC in the signal chain before the detectors (for ring artifacts)
  • Noise may be normal at high-resolution scan settings
Checks, in order
  1. Inspect the fibre-optic lines and cables
  2. Repeat Air and NRA calibrations
  3. Check the detector supply rails with an oscilloscope (one case measured 50 mV ripple against a 100 mV tolerance)
  4. Scan the image quality phantom with the settings given in the maintenance manual and compare the standard deviation with the tolerance
Fix

All three cases are probable. For interference that starts mid-scan, find and repair the damaged fibre-optic cable. For persistent ring artifacts, colleagues recommended replacing the detectors first, with the detector electronics as an alternative suspect. For noise, run the phantom image quality check: the image may well be within tolerance.

Storage arrays fail or will not rebuild with replacement disks. On the RAW data disk (RDD) array, mixed disks would not work together. On the IDD server, ordinary SAS/SATA/SSD replacement disks were not recognised and a 5-disk RAID would not rebuild. On another system, crashes appeared to come from the RAID.field case
Likely causes
  • Incompatible disk types mixed in one RDD RAID array
  • Wrong disk firmware revision
  • The server needs a non-standard sector size and a specific disk firmware
Checks, in order
  1. Review the logs and the RAID array status
  2. Confirm the disk type, interface and firmware revision before mixing disks
  3. On the IDD server, check the disk revision and sector format
Fix

RDD (Aquilion 64), solved: use compatible Fibre Channel disks with firmware 0003. The 73 GB (BSX73-2232E) and 300 GB (BSX73-2396E) disks can be mixed in one array; the 36 GB disk (BSX73-1828E) cannot. IDD server, solved in one engineer's experience: an equivalent standard disk worked after being reformatted to 520 bytes per sector under Linux. In the RAID-crash case (probable), the disks were repaired or replaced. Keep the software media and manuals on site.

Console software faults. Cases seen: after patient registration no scan protocols can be selected, although warm-up and calibrations still work. The D console hangs at a grey 'Start-up failed' message, then shows a black screen. A workstation shows 'fsys can not run' and the application shows no patients, although the disks and partitions read fine from Windows. Voice commands do not record. Transfers between consoles stop on contrast studies, with HOLD shown next to the remaining series.field case
Likely causes
  • Corrupted application software
  • Damaged system disk on the D console
  • Consoles cannot see each other on the internal network
  • Badly installed sound card drivers
  • Faulty reconstruction unit (for the HOLD transfer problem)
Checks, in order
  1. Confirm the reconstruction subsystem is fully booted (recon, boards, RDD and GCIFA all ready)
  2. Check the HDD and memory, and run chkdsk and defragmentation
  3. Check the internal network settings, the hub in the CPU, and the console network cards
  4. Check the disks for errors and confirm the configuration files are readable
Fix

Solved: protocol selection came back after reinstalling the software (restoring a saved backup is an alternative), and voice commands came back after reinstalling the sound card drivers. Probable: for the D console, clone the disk from the S console and reconfigure it following the installation manual. For 'fsys can not run', check the inter-console network first. For the HOLD transfers, suspect the reconstruction unit.

Network and DICOM problems. Cases seen: studies do not reach PACS even though ping and telnet to the PACS succeed. A new workstation needs to be added as a transfer destination. In another case, all links between the CT units were lost during installation or service.field case
Likely causes
  • Fault on the PACS server side
  • DICOM node not configured
  • Lost node configuration
Checks, in order
  1. Run ping and a DICOM echo from the CT to the PACS
  2. Send a test study to a different destination to isolate the fault
  3. Check the disks and the DVD drive (in one case a disc was stuck in the drive)
  4. Photograph the node configuration before any service work
Fix

Probable: the PACS-side fault was likely in the PACS server itself, so set up a second destination for testing. To add a workstation, register it as a DICOM node in the service settings and reboot; it then appears in the transfer list. Solved: the lost node links were re-entered from a photo of the previous configuration.

Tube problems. With heavy contrast workloads, the scanner blocks scanning because the tube is too hot, once pausing a study for about 15 minutes. In another case, the gantry shook hard during rotation spin-up after a tube replacement.field case
Likely causes
  • Many back-to-back contrast studies with poor room cooling (office-grade air conditioners)
  • After tube replacement: vibration from the tube insert, or a gantry imbalance
Checks, in order
  1. Monitor the tube heat percentage
  2. Review the daily number of contrast studies and the room cooling
  3. After tube replacement, find whether the vibration comes from the tube insert or from gantry balance, and confirm the tube is original
Fix

Probable. For overheating: avoid scheduling contrast studies back-to-back, let the tube cool (keeping it at about 40% heat or below), and reduce protocol phases where clinically acceptable. For vibration: if it comes from the tube insert, the insert must be scrapped. If it is a gantry balance problem, use the gantry balancing procedure (there is no separate rotor balancing procedure). Do not run the gantry at speed with abnormal vibration.

05

Field notes

Calibration order (solved case): do the Air calibrations first. Then run NRA starting with the LL phantom, selecting all voltages and the full table, and follow the system prompts. Repeat the Air calibrations after NRA.

Don't overlook the small 24 V power supply on the HV multiplier: in two Aquilion 64 cases it caused 'over kV' errors, and one showed up only at high mA.

If the tube temperature reads 100% at power-up with an XC error, check the timekeeper battery. After replacing the timekeeper, recalibrate the filament currents.

Table servo faults are often mechanical or encoder-related. Clean the encoders, check their bearings, and look for disagreement between the motor encoder and the gearbox-output encoder before replacing electronics.

Take backups and photographs of the console configuration and node links before service work. Several recoveries depended on a backup, a photo, or cloning the twin console.

Disks in the RAW and image storage arrays are not generic. Match the type, interface, firmware revision and sector format before inserting a replacement.

After any repair, run the manufacturer's full checkout and performance verification before the device goes back to patients.