A-dec 500 (300): broken chair backrest side shells
What happened
Both side sections of the chair backrest had snapped off and were held in place only by the upholstery. As reported in the case, the plastic was considered beyond normal repair and a replacement part was being sought.
What was checked
- Assessed whether the plastic could be repaired at all — as reported in the case, ordinary adhesives do not bond to this material.
- Looked for a donor backrest in parallel, in case the repair was not feasible.
Cause
Mechanical overload: as reported in the case, patients pushing themselves up while leaning on the armrest load the thin side sections of the backrest shell until they break off.
What was done
Since the plastic does not take ordinary glue, it was welded: a temperature-controlled soldering iron with a heat gun, filler pieces cut from the same plastic, and short pins of strong wire melted in across the cracks to spread the load. The alternative reported in the case was a fast-curing bumper repair compound (epoxy with fiberglass), laid in several layers over the prepared seam with an overlap of about 4–5 cm as reported in the case. Both methods gave a strong, durable joint.
Safety
Shield the upholstery, tubing and wiring from the heat gun and molten plastic before you start, and do not return the chair to use until the repaired shell can clearly carry patient load — otherwise replace the part.
Lesson: When a plastic part will not take glue, welding it with filler from the same material plus wire reinforcement can save a part with no ready replacement — but only if the repair can be trusted to carry the load.
Covered in the course: Dental Equipment Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.