Carestream CS2200: exposure error on detailed imaging

What happened

On detailed exposures the unit aborted with error E15 (E13 was also mentioned, as reported in the case). The manual points to a kV feedback exceedance, and with a breakdown the kV sags, so the first suspicion was arcing inside the tube.

What was checked

Cause

Not a tube breakdown: the fault was in the cable at the monoblock-side connector, where a damaged conductor/connection disturbed the feedback and triggered the error, as found in this one case.

What was done

The engineer repaired the faulty cable and connection at the monoblock-side connector, after which exposures completed without the error, as reported in the case.

Safety

Disconnect the unit from the mains and follow the manufacturer's discharge procedure before touching high-voltage connectors, and after any cable repair confirm electrical safety and perform test exposures before returning the unit to clinical use.

Lesson: An error that reads like a tube breakdown can come from the cable and its connectors, so inspect the whole high-voltage path before condemning the tube.

Covered in the course: Dental Equipment Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.