Csn RX4: intraoral sensor intermittently drops off USB
What happened
The sensor was intermittently detected by the computer as an unknown USB device: it would start working, then drop off again — with no clear link to how the cable was bent.
What was checked
- Reseated the internal connector and rang out the cable — continuity checked out fine.
- Tried the sensor on different computers and without the USB extension to rule out power problems.
- On the board, the engineers found an Altera controller and a Cypress CY7C68013A USB controller; the dual choke in the USB data pair was also on the suspect list.
Cause
A solder defect — cold/cracked joints on the Cypress CY7C68013A USB controller — intermittently broke the USB link, which is why the fault did not follow cable movement.
What was done
Board-level rework: a specialist reballed (re-flowed) the Cypress CY7C68013A USB controller without overheating the assembly. The sensor matrix survived the soldering, and the unit was reported working normally afterwards. This is precision BGA work — not a job for a standard soldering iron.
Safety
Before the sensor goes back into service, repeat the electrical safety checks (protective earth continuity and leakage current) and disinfect it as an intraoral device, per protocol.
Lesson: When a USB device drops out with no dependence on cable movement, suspect a cracked solder joint on the board's USB controller, not just the cable and connectors.