Csn RX4: intraoral sensor intermittently drops off USB

What happened

The sensor was intermittently detected by the computer as an unknown USB device: it would start working, then drop off again — with no clear link to how the cable was bent.

What was checked

Cause

A solder defect — cold/cracked joints on the Cypress CY7C68013A USB controller — intermittently broke the USB link, which is why the fault did not follow cable movement.

What was done

Board-level rework: a specialist reballed (re-flowed) the Cypress CY7C68013A USB controller without overheating the assembly. The sensor matrix survived the soldering, and the unit was reported working normally afterwards. This is precision BGA work — not a job for a standard soldering iron.

Safety

Before the sensor goes back into service, repeat the electrical safety checks (protective earth continuity and leakage current) and disinfect it as an intraoral device, per protocol.

Lesson: When a USB device drops out with no dependence on cable movement, suspect a cracked solder joint on the board's USB controller, not just the cable and connectors.