Fona 1000: control buttons unresponsive, chair won't move despite beeping

What happened

On a dental unit the control buttons stopped doing anything: the unit beeped as if powered and working, but no motor would run, so it looked like it had locked itself into a protection state.

What was checked

Cause

The motors are fed from their own supply line, separate from the one powering the control panel and beeper. Its fuse had blown, so the unit beeped and accepted button presses but could not move anything — it only looked like a protection lockout, while it was a plain power fault.

What was done

The engineers replaced the blown fuse with one of the same type and rating, and the motor functions came back; the outcome in the case was reported as solved.

Safety

Isolate the unit from the mains before opening it, verify with a meter that the internals are actually de-energised, never fit a fuse with a higher rating or a makeshift substitute, and run an electrical safety test before the chair returns to clinical use.

Lesson: A beeping panel only proves that one supply branch is alive — before calling a fault a protection lockout, check whether the drives have their own feed and their own fuse.