Konica Minolta · 873 (DryPro 873)

Konica Minolta DryPro 873: film size will not change

What happened

The imager would not let the engineers change the film size to 8×10, neither at the device nor in the connected software, even though the tray had been reloaded with the correct size.

What was checked

Cause

A job for the old tray/format was stuck in the print queue, and the printer blocked any film size change while that job was pending.

What was done

The engineers loaded a different film format so the printer could output the stuck job; once the queue was clear, the size changed normally both at the printer and in the connected imaging software.

Safety

Keep the light-tight covers and the laser exposure unit closed while the imager is powered, and confirm all panels and interlocks are back in place before returning it to use.

Lesson: If a film size will not change after reloading a tray, clear or complete pending print jobs first — a queued job for the old format can lock the size setting.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.