Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Ranyes Valencia
Ranyes · Valencia

Ranyes Valencia: dental chair will not lower fully

What happened

Reported in the case: a Ranyes Valencia dental chair stopped part-way and would not complete its travel to the fully lowered position.

What was checked

Cause

The engineers found that the memory chip (DIP-8) on the chair control board had failed, so the board no longer held the taught limit positions and the chair would not go all the way down.

What was done

The engineers replaced the DIP-8 memory chip on the chair control board, then ran the position teach-in procedure from the manufacturer's service documentation, done by authorised personnel. Reported in the case, full downward travel was restored after the replacement and re-teaching.

Safety

Before opening the chair control box, isolate the chair from the mains and confirm the supply is dead, and when testing the full travel keep hands and feet clear of the moving base, making sure nothing and no one is under the seat.

Lesson: When a chair will not accept or keeps losing its taught limit positions, suspect the memory chip on the control board before blaming the adjustment routine itself.

Covered in the course: Dental Equipment Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes