Ranyes Valencia: dental chair will not lower fully
What happened
Reported in the case: a Ranyes Valencia dental chair stopped part-way and would not complete its travel to the fully lowered position.
What was checked
- First, following the usual routine for this class of chairs, the engineers checked the programmed positions of the upper and lower limits, the same way it is done on similar Chinese-made chairs.
- When the travel still would not complete, attention moved to the chair control board and its DIP-8 memory chip.
Cause
The engineers found that the memory chip (DIP-8) on the chair control board had failed, so the board no longer held the taught limit positions and the chair would not go all the way down.
What was done
The engineers replaced the DIP-8 memory chip on the chair control board, then ran the position teach-in procedure from the manufacturer's service documentation, done by authorised personnel. Reported in the case, full downward travel was restored after the replacement and re-teaching.
Safety
Before opening the chair control box, isolate the chair from the mains and confirm the supply is dead, and when testing the full travel keep hands and feet clear of the moving base, making sure nothing and no one is under the seat.
Lesson: When a chair will not accept or keeps losing its taught limit positions, suspect the memory chip on the control board before blaming the adjustment routine itself.