UZT 1.07 (digital): diagnostic short takes out the MCU board
What happened
The fault initially called for replacing a single faulty transistor, but during diagnostics the microcontroller board was accidentally shorted, turning a minor component swap into major board damage.
What was checked
- Short-circuit diagnostics on the board: the engineers found the microcontroller board shorted during their checks.
- Continuity measurements across the board to confirm which components, including the microcontrollers, were affected by the short.
Cause
The original fault was one bad transistor; the microcontroller board damage came from a probing slip during diagnostics, not from the equipment itself.
What was done
The engineers replaced the microcontrollers in hardware: they wrote replacement firmware for two ATmega MCUs and fitted them on the board. Per the case report, the unit has run without repeat faults for three years; note that this is a firmware-level workaround, not a manufacturer-documented repair path.
Safety
Isolate the unit from the mains and discharge the filter capacitors before probing any board, and run the full electrical safety test before returning the device to use.
Lesson: A single careless probe move can turn a one-transistor repair into a full board rebuild, so protect the board during diagnostics as carefully as you repair it.
Covered in the course: Dental Equipment Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.