UZT 1.07 (digital): diagnostic short takes out the MCU board

What happened

The fault initially called for replacing a single faulty transistor, but during diagnostics the microcontroller board was accidentally shorted, turning a minor component swap into major board damage.

What was checked

Cause

The original fault was one bad transistor; the microcontroller board damage came from a probing slip during diagnostics, not from the equipment itself.

What was done

The engineers replaced the microcontrollers in hardware: they wrote replacement firmware for two ATmega MCUs and fitted them on the board. Per the case report, the unit has run without repeat faults for three years; note that this is a firmware-level workaround, not a manufacturer-documented repair path.

Safety

Isolate the unit from the mains and discharge the filter capacitors before probing any board, and run the full electrical safety test before returning the device to use.

Lesson: A single careless probe move can turn a one-transistor repair into a full board rebuild, so protect the board during diagnostics as carefully as you repair it.