01Lesson 1 / 5 · 23 min

The Primus platform: modules, boards and the gas path

Identify the Primus configuration and software in front of you, name its modules, boards and pneumatic components, and follow the gas from the pipeline inlet through the electronic mixer, vaporiser, compact breathing system and piston ventilator to the scavenger, knowing where it differs from a Fabius.

Four modules on one bus

The Primus is easiest to understand as a small network of modules rather than as one machine. The user interface, called the GUI, is made of the monitor control panel (MoBi) with its 12-inch colour screen, membrane keys and central rotary knob, the S-Box with the external interfaces and the optional SpO2 module, and the patient gas module (PGM). Next to it sit the electronic mixer, the VGC (ventilation and gas controller, which is the ventilator plus the compact breathing system) and the power pack with its UPS batteries. Optional cylinder pressure reducers hang off the gas inlet block. The modules exchange settings and status over a CAN bus, and each one has its own processor and its own software version.

All the harnesses meet on one board, the NEUTRAL POINT PCB. It distributes the supply voltages, the CAN lines and a few discrete signals: the main switch, the microswitch in the safety O2 adjuster, the lamp, the PGM and the connectors for MoBi, mixer, VGC and power pack. Remember that board. When one module stops talking to the others, the cause is as often a connector or harness at the neutral point as a failed processor board.

This course follows the Primus Technical Documentation, revision 6.0, and the test list bound into it. The documentation describes two software generations. Up to software 1.n the modes are IPPV, SIMV, PCV and MAN/SPONT; from 2.n they are renamed Volume Mode, Pressure Mode, optional Pressure Support and Man/Spont, with switchable synchronisation. Several service details change with the software, as later lessons show, so read and record the version of every module: MoBi, mixer master and supervisor, VGC master and supervisor, PGM and power pack. Later members of the family and newer software have their own documentation; where they differ, theirs wins.

Then note the options, because they decide which tests apply: PGM or PGM2, SpO2, the external fresh gas outlet (A-cone), cylinders with high-pressure reducers, a bronchial suction device, an external O2 flowmeter, one or two vaporiser positions and which APL valve design the breathing system has.

If you know the Fabius, unlearn this first

Engineers who come from the Fabius find a familiar piston and a familiar compact breathing system, and then several surprises. The Primus has no mechanical flow control valves and no flow tubes for the fresh gas. The mixer is electronic: valves meter each gas into a small tank, and a proportional valve delivers the set flow. There is no pneumatic ratio controller to adjust; the minimum O2 rule is enforced by the mixer software. The backup for a mixer or system failure is a separate mechanical safety O2 adjuster that feeds O2 straight to the vaporiser.

Gas monitoring is built in. The PGM samples from the Y-piece and measures O2, CO2, N2O and agents, so a measurement complaint can be a pneumatic fault inside the machine. The ventilator piston needs vacuum on its lower rolling seal before the motor is allowed to move, as on the Fabius, but on the Primus the pump, the PEEP valve and the switching valves sit in a VGC pneumatic block inside the ventilator drawer. The breathing system is heated, has two electrical flow sensors and a fresh gas isolation valve. And the power pack is a separate module with two 12 V lead-gel batteries that run the whole workstation, ventilator included, through a mains failure.

Supplies and the gas inlet block

The pipeline gases enter at the gas inlet block on the underside of the mixer through NIST or DISS connectors. The documentation expects pipeline pressure between 270 and 690 kPa. If it falls below 270 kPa the user is expected to open the cylinder, and the maximum cylinder pressure the machine accepts is 19,000 kPa. For each pipeline gas the path is the same: a sintered-metal filter that protects what follows, a relative pressure sensor that tells the mixer the supply pressure, a non-return valve that stops cylinder gas escaping into a failed pipeline, and a pressure regulator (PRPO2, PRPAIR, PRPN2O) that feeds the mixer at 2.3 bar plus or minus 0.1 bar. The O2 regulator also supplies the flush button and the safety O2 adjuster, which is why a weak O2 regulator shows up in three places at once.

Cylinders are optional. Each cylinder has its own reducer (from 200 bar for O2 and air, from 60 bar for N2O) set by the manufacturer to an outlet of 5 to 6 bar, or 2.7 to 3.5 bar on the US version, with an internal blow-off valve that limits the outlet if the valve cone fails to close. High-pressure sensors report the cylinder contents to the mixer.

The supply status is shown by LEDs on the front panel, and here the Primus helps you. Even with the main switch off, as long as the mains plug is in, the mixer runs in a low-energy state and reads the supply sensors, so the LEDs tell you what is connected before you start the machine. A pipeline LED is green from 2.7 bar; a cylinder LED is green from 20 bar for O2 and air and from 10 bar for N2O. A flashing green pipeline LED means a sensor error, not a low supply. A flashing red cylinder LED means the cylinder is low and the matching pipeline has failed too: the dangerous combination.

The electronic mixer

The mixer is the MIXER PCB, the PRIMUS CPU PCB, the gas inlet block, the mixer block, a 0.5 L fresh gas tank and the optional A-cone valve. The user sets the carrier gas (air or N2O), the total fresh gas flow and the O2 concentration. The inlet valves VMIXO2, VMIXAIR and VMIXN2O open in turn and fill the tank between about 1 and 1.5 bar. A differential pressure sensor across a flow restrictor (PDMIX with RM) measures how much of each gas went in, so the concentration is built up by volume, filling by filling. The tank pressure sensor PTANK watches the tank, corrects for ambient pressure and cross-checks the flow measurement; temperature sensors in the tank and block correct for gas temperature. At the full 18 L/min the inlet valves cycle about 72 times a minute.

On the outlet side the proportional valve VMGS delivers 0.2 to 18 L/min, regulated by a second flow measurement (PDMGSHI and PDMGSLO across RMGS) with PSYS compensating for ambient pressure. VTANK vents the tank when both O2 and air are missing. The bistable A-cone valve VSWAK switches fresh gas to the external outlet and reports its position; a mechanical safety valve, VBAK, limits that outlet. The flush button is purely mechanical and is specified at a minimum of 35 L/min.

The PRIMUS CPU PCB has two identical processors, a master and a supervisor, each with its own watchdog and configuration memory, and 7-segment displays that show their states. The supervisor checks the current in every inlet valve through a shunt; the proportional VMGS cannot be checked that way. If something is wrong, the mixer can cut the supply to all its valves. Two start-up rules matter for service. With the plug in and the switch off, only a limited-energy supply reaches the mixer interior, because a small internal leak could have enriched the inside of the mixer with O2. At switch-on the mixer fan must run for at least 10 seconds before the other modules get power, and the fan speed signal is monitored.

The VGC: piston, rolling seal and vacuum

The VGC drawer holds the piston-cylinder unit (PCU), the VGC pneumatic block, three boards and the interface plate that carries the breathing system. It receives the unstabilised 18 to 30 V supply, 12 V for its fan at once, 12 V for the small pneutronic valves about 10 seconds later, and the CAN bus, and it makes its own 5, 15 and 24 V. The VGC POWER PCB drives the piston motor, the vacuum pump, the valves and the breathing system heater, which it watches with two independent NTCs. The PRIMUS ANALOG PCB drives the PEEP valve, reads the two airway pressure sensors, the vacuum sensor and the flow sensors, counts the motor encoder and reads two light barriers: one for the piston end position and one that confirms the drawer is fully inserted. A third board, a PRIMUS CPU PCB like the mixer's, runs the ventilator.

In the PCU the motor turns a nut on a spindle fixed to the piston. One turn moves 38.5 mL, so the full 1400 mL stroke is about 37 turns, and the encoder gives 1024 pulses per turn. At start-up the piston is driven down to the light barrier to set its reference. The piston is sealed by a loose lower rolling seal held in shape by vacuum, and topped by the upper bag rolling diaphragm. The motor is only released when the pump has built 120 to 250 mbar of vacuum. The pump draws through a bacterial filter; a valve set to about 200 mbar, two buffer volumes and a restrictor shape the vacuum; RV3 holds it in the diaphragm space when the pump stops in MAN/SPONT so that the return to automatic ventilation is quick. MV1 is the PEEP/Pmax valve, MV2 the calibration and vent valve, MV3 switches pressure measurement between breathing system and A-cone, MV4 is the PEEP control valve.

The ventilator covers tidal volumes from 20 mL (10 mL from software 2.n) to 1400 mL at 3 to 80 breaths per minute with peak flows up to 150 L/min, all without drive gas.

Breathing system, gas monitoring and the way out

The compact breathing system sits on the interface plate and is locked by three screws. Inside are the inspiratory and expiratory valves with ceramic discs, the fresh gas isolation valve RV1, the PEEP diaphragm valve V1, the APL bypass valve V2, the AGSS non-return valve RV2, the APL valve, the absorber, a heater in the valve plate and two electrical flow sensors, one after the inspiratory socket and one at the expiratory socket. RV1 closes during mechanical inspiration so that only the small volume between RV1, the inspiratory valve and V1 adds compliance; fresh gas collects in the bag meanwhile. In automatic modes V2 is held open, so the APL setting has no effect, and V1 doubles as the Pmax limit, adjustable from 20 to 70 hPa. In MAN the APL limits pressure between 5 and 70 hPa; in SPONT it is open. Two APL designs exist with different covers on the same valve plate and block. Surplus gas leaves through RV2 to the scavenger.

The PGM (IRIA bench with an electrochemical O2 cell) and the PGM2 (ILCA2 bench with a paramagnetic O2 sensor) draw sample gas through a water trap with two hydrophobic membranes, a Nafion drying tube, a zeroing valve that switches to room air, the analysers, a pneumatic low-pass filter and the pump, then return it to the breathing system. The PGM pumps about 150 mL/min, the PGM2 about 200. A PGM2 has no O2 cell holder on the rear; its type plate sits there instead.

So the whole path reads: pipeline or cylinder, inlet filter and sensor, non-return valve, regulator, mixer tank and proportional valve, vaporiser on the plug-in mount (one at a time by interlock), fresh gas inlet of the breathing system or the A-cone, piston and patient, and out through RV2 to the AGSS.

Unit breakdownPrimus main assemblies
PartWhat it doesHow it fails
01NEUTRAL POINT PCBCentral distribution of supply voltages, CAN bus and discrete signalsLoose connectors or harness faults that look like module failures
02MoBi (monitor control panel)Screen, keys, knob; two processors running GUI and PGM evaluationDisplay or backlight faults, keypad wear, lost settings if buffering fails
03Gas inlet blockFilters, supply sensors, non-return valves, regulators to 2.3 barBlocked filter, leaking non-return valve, regulator drift, sensor error
04Mixer block and fresh gas tankInlet valves, flow and pressure sensors, proportional fresh gas valveValve or sensor faults reported by the supervisor; mixer to safety mode
05PRIMUS CPU PCBMaster and supervisor processors in the mixer and in the VGCWatchdog or memory faults; early software used a lithium cell
06Safety O2 adjuster and flushMechanical O2 backup flow and mechanical flush from the O2 regulatorLeft open (blocks the self-test), sticking flush button
07Piston-cylinder unitMotor, spindle, encoder, light barrier, rolling sealsRolling seal damage, encoder disc bent, reference not found
08VGC pneumatic blockVacuum pump, PEEP/Pmax valve, switching and vent valves, filterVacuum not reached, blocked filter, valve faults
09Compact breathing systemValves with ceramic discs, RV1, V1, V2, RV2, APL, absorber, heater, flow sensorsDirt on valve seats, worn diaphragms and O-rings, seal leaks
10PGM or PGM2Side-stream O2, CO2, N2O and agent measurement with water trap and pumpWet membranes, kinked or tired Nafion tube, pump flow out of range
11Power pack with UPSMains to 24 V and 12 V outputs; two 12 V lead-gel batteriesInput stage damage, aged batteries, filter mat clogged
Fault-finding flowSupply LED off, flashing or disagreeing
A pipeline is connected, butits LED stays off or flashes,or the checklist shows apressure that disagrees withyour gauge.Does a test gauge atthe wall outlet readat least 2.7 bar?1NoLow or no supply: the hospitalsystem or the hose, not the machine.YesIs the hose probefully engaged in itsNIST or DISS inlet?2NoReseat it; inspect the inletsintered filter for blockage.YesIs the LED steadygreen, not flashinggreen?3NoFlashing green is a sensor error:sensor, harness or MIXER PCB.YesIs the checklistpressure within 200mbar of your gauge?4NoThat pipeline sensor or itsamplifier is out of tolerance.YesDoes the fresh gasflow test reach itstargets on that gas?5NoLook at the non-return valve, filteror regulator of that inlet.YesSupply path proven: record thereadings and look for thecomplaint elsewhere.
Trainer: walk the fault-finding flowAnswer each check as if you were at the machine. The flow leads you to the cause.
field case

Oil in the air line

A hospital replaced the dedicated compressor of its medical air plant with an old oil-lubricated unit that had no dryer. Water mixed with oil travelled down the pipeline into a Primus. During a test, oil vapour in the breathing system ignited, apparently at the heated element of a flow sensor, and the inspiratory and expiratory valves and the circuit were destroyed. Colleagues who discussed the event described how they recover a machine after milder contamination: dismantle and dry the parts of the gas path, dry the inlet filter, blow the lines through and let clean gas run through the machine for hours before testing.

Gas quality is part of the gas path. Oil and water from a poor air plant reach every component downstream of the inlet, and on this machine there are heated elements in the gas stream. A machine contaminated with oil must not be returned on the strength of drying alone: contaminated parts are replaced, the source is fixed, and the full checkout is done.

From service practice

A regulator from another Dräger machine

An engineer asked colleagues whether the gas inlet pressure regulators from another Dräger ventilator or from a Fabius would fit a Primus, since they looked similar and one was available. The only practical advice in the discussion was to compare the order numbers in the parts catalogues. That is the right instinct. On the Primus the regulators in the gas inlet block feed the electronic mixer at a defined inlet pressure, carry a flow direction marking and are fitted with specific O-rings, and the mixer's metering assumes that pressure.

Similar-looking parts from a related machine are not interchangeable until the parts catalogue for this machine says so. Fit the specified regulator in its marked flow direction and verify its output and the fresh gas accuracy afterwards.

On-site checklist

  • Data plate and serial number recorded
  • Software version of MoBi, mixer master and supervisor, VGC master and supervisor, PGM and power pack recorded
  • PGM or PGM2 identified (O2 cell holder on the rear or type plate)
  • Options noted: SpO2, A-cone, cylinders and reducers, suction device, external O2 flowmeter
  • Vaporiser positions and interlock type noted
  • APL valve design and breathing system serial numbers noted
  • Supply LEDs read with the plug in and the switch off

Check yourself

1. Mains fails during a case on a Primus whose UPS batteries are healthy and charged. What should the users expect?
2. With the plug in and the main switch off, the O2 pipeline LED is flashing green. What does that tell you?
3. An engineer who knows the Fabius looks for the oxygen ratio controller to adjust on a Primus. What should a colleague tell him?
4. The ventilator refuses to start and the service screen shows that the vacuum on the rolling seal stays at 60 mbar. Why does that stop the piston?
5. Which statement about fresh gas on the Primus is correct?