Early access: courses are being reviewed by field engineers
Service EngineerAcademy
01Lesson 1 / 5 · 22 min

The Vivid platform: Vivid 7 and Vivid i side by side, and the signal path of each

Identify which Vivid you are facing and at what build level, name the main assemblies of the Vivid 7 and the Vivid i and their place in the signal and power path, and know where the two machines differ from each other and from a LOGIQ before you open either.

Two machines, one cardiac family

This course covers two members of GE's cardiovascular Vivid family that you still meet in the field. The Vivid 7 is the large cart system of the early 2000s; its service manual (FC091194, revision 02, 2002) covers both the Vivid 7 and the Vivid 7 PRO at two hardware build levels, BT'01 and BT'02. The Vivid i is the laptop-style portable; its manual (2378956-100, revision 5, 2007) describes a unit of about 5 kg that GE presented as carrying the same imaging technology as the larger Vivid systems into a portable case, optionally mounted on a SafeLock cart.

Both are phased and linear array scanners built first for cardiology: sector imaging with phased arrays, ECG-triggered acquisition, PW and CW Doppler, colour and tissue Doppler, transoesophageal (TEE) probes and stress echo. The generic physics, the phantom and the probe passport were covered in Ultrasound Service — Level 1, the general fault-finding method in Level 2 and probe testing in Ultrasound Probes. Here we add only what these two machines do differently and how they fail.

If you come from LOGIQ, three things change. First, the clinical workload is cardiac, so ECG, TEE and CW matter as much as B-mode. Second, the Vivid 7 front end is not one master board but a rack of separate boards, each of which can be tested and replaced on its own. Third, the Vivid i is not a LOGIQ e in a different case: its front end, power supply and battery are its own design, with their own tests.

First job on site: establish exactly what you have

On a Vivid 7, read the rating label and the part number. The manual lists separate part numbers for 230 V and 100-120 V versions of the Vivid 7 and the Vivid 7 PRO, and for the BT'01 and BT'02 build levels. Build level matters: the manual carries a table of required board revisions for each build level and software generation (software 1.x on BT'01, 2.x on BT'02), and it warns that the normal software reload procedure is not the way to upgrade a BT'01 unit to BT'02. Note also which back end processor is fitted. The manual describes two generations, BEP-1 and BEP-2, with different PC hardware and different system software discs.

On a Vivid i, read the catalogue number on the base, the software version and whether the unit lives on a SafeLock cart. Then, on either machine, record the probes and their serial numbers, the installed options, the peripherals and the network connection. Photograph the configuration screens. This is your baseline for every later decision: which spare part revision fits, which software disc is valid and which options the customer should still have after a repair.

Inside the Vivid 7: card rack, back end and power

The manual divides the Vivid 7 into a front end processor (the card rack), a back end processor, the top console, power distribution, internal and external I/O, peripherals and a modem. The signal runs from the probe connector panel through the front end and the mid processors into the back end and on to the monitor and peripherals. Configuration lives on the hard disk and all software loads from it at power-up.

The front end proper has five board types. The relay board (RLY) carries the three probe connectors and selects one of them through relays. The transmitter (TX128) generates the delayed transmit pulses; their amplitude comes from the TX power module and is set by the acoustic power software. The receiver (RX128) has transmit/receive switches at its input, then preamplification and analogue TGC. Two beamformer boards (BF64) each digitise 64 channels and form the beam. The front end controller (FEC) runs the whole rack: it loads scan parameters into TX and beamformer memory, reads the probe identification, selects the connector, drives the high voltage multiplexer in large linear probes and generates the system clock. Two transducer bus boards (XDBUS) carry signals between the RLY board and the TX and RX boards.

The mid processors follow: the RF and tissue processor (RFT), the spectrum Doppler processor (SDP) and the image port (IMP2), linked by a one-way data bus that GE calls the Pipelink. The back end is a PC in a shielded enclosure with its own ATX supply and a UPS battery, a GE-designed PCI card (PC2IP) that is the link to the front end, a hard disk, a CD-RW drive and an optional magneto-optical drive. The physio module that handles ECG and other analogue inputs is part of the back end.

Power enters through an EMI filter and a mains switch that trips on overcurrent, passes an inrush limiter and feeds an isolating transformer, which is the galvanic barrier between the site mains and the scanner. The transformer comes in a 230 V and a 115 V version. From it, the AC distribution feeds the card rack, the back end, the monitor, the modem, the peripheral outlet (selectable 115 or 230 V) and the fan supply. In the card rack a switch-mode DC power module supplies the boards through the backplane, and a separate TX power module supplies the programmable transmit voltages and the multiplexer supply for large linear probes.

Inside the Vivid i: three blocks in a laptop

The Vivid i manual divides the electronics into three: front end, back end processor and power supply unit, plus the keyboard with the ECG module and the peripherals. The front end is three boards. The radio frequency interface (RFI) board combines the functions that the Vivid 7 spreads over the FEC, RFT and IMP boards. Two identical TR32 boards each carry 32 transmit and receive channels with their own digital beamformer, 64 channels in total. The Probe and MUX board takes the 128 channels from the probe connector, selects 64 of them with a high voltage multiplexer, and also holds the analogue CW Doppler processor and the transmitter and preamplifier for the pencil probe.

The back end is an ETX processor module on a base board, with a hard disk, an LCD interface for the 15-inch panel, PCMCIA (CardBus) slots, USB, Bluetooth and fan control. A protection plug on the base board holds the system's software options; the manual states that the system will not operate properly without it. The power supply unit selects between the external adapter and the internal battery, generates the low voltages for the back end and front end, generates the high voltages for the transmitters and multiplexers, charges the battery and reports to the processor over the SMBus. The battery is a lithium-ion pack of 12 cells. Five fans pull warm air out through the side vents. Mains power never enters the case: an external AC-DC adapter accepts 100-240 V AC and delivers 20 V DC to the DC input on the rear panel.

Probe connection is simpler than on the cart: one array probe port with a locking lever and one pencil probe connector, both on the right side. The ECG module sits in the keyboard assembly and supports three bipolar leads (I, II, III) plus an external ECG input.

Cardiac features and what they mean for service

Every cardiac feature adds something to check. ECG inputs are type CF applied parts with much tighter leakage limits than the probes, so any repair near the physio module or, on the Vivid i, the keyboard, ends with an ECG leakage test. TEE probes have their own manuals and their own leakage test schedule, and a cracked or damaged TEE probe is a patient risk, not a cosmetic defect. CW Doppler uses a dedicated pencil probe connector on both machines, so a CW complaint may be the pencil probe, its connector or the CW circuitry, not the main probe path.

Options such as tissue velocity imaging, contrast, stress echo and multi-image are software options bound to the system. The Vivid 7 warns at start-up when installed options, typically demo options, are about to expire or have expired. Record the option list before any repair, and never promise a customer that options will come back by themselves after a disk or back end change.

  • Vivid 7: cart, mains inside, isolating transformer, card rack of separate boards, three array connectors on the relay board plus a pencil connector, PC back end with UPS battery, 17-inch monitor, rear air filter.
  • Vivid i: portable, external 20 V adapter, three-board front end with 64 channels, one array port plus pencil port, ETX back end, lithium-ion battery, five fans, optional SafeLock cart with its own AC unit.
  • Both: up to 192-element linear probes through multiplexing, configuration and software loaded from the hard disk, a common GE service platform with logs and diagnostics.

Where the two machines differ, and where both differ from LOGIQ

On the Vivid 7 you think in boards: a fault can be pinned to RLY, TX, RX, BF, FEC or a mid processor with the built-in tests, and those boards are individual field-replaceable units whose revision must match the build level. On the Vivid i you think in three larger assemblies: Probe and MUX, TR32 and RFI on the front side, BEP and power supply behind them. Power faults also look different. A Vivid 7 lives on the mains with a small UPS that only lets the PC shut down cleanly; a Vivid i runs from adapter or battery, and its battery, charger and fans are part of daily fault-finding.

Compared with a LOGIQ e, the Vivid i has a separate high voltage supply that the manual describes in detail, a CW pencil probe path, an ECG module in the keyboard and five fans instead of two. Compared with cart LOGIQs, the Vivid 7 has a fully separate front end rack with its own diagnostics and calibration. What carries over is the method: identify, read the logs, separate probe from port from front end from back end, back up before software, and verify after repair. Board names, voltages and procedures come from each machine's own manual.

Unit breakdownVivid 7 and Vivid i main assemblies
PartWhat it doesHow it fails
01Vivid 7 relay board (RLY) and XDBUS boardsThree probe connectors, relay selection, signal routing to TX and RXDead or noisy single port, bent pins, relay that does not switch
02Vivid 7 TX128 and TX power moduleTransmit pulse generation and programmable transmit voltagesWeak or missing image on all probes, TX power errors in tests
03Vivid 7 RX128 and two BF64 boardsT/R switches, preamp, analogue TGC; digitising and beamformingChannel dropouts, stripes, failed DC offset calibration naming a BF board
04Vivid 7 FEC, RFT, SDP, IMP2Front end control and clocks; tissue, Doppler and image data pathsNo image in one mode, Test All failures on the named board
05Vivid 7 back end processor with UPS batteryPC: software, user interface, archive, PC2IP link to the front endNo boot or beeping after long power-off, disk errors, no front end link
06Vivid 7 AC controller, transformer and DC powerMains entry, isolation, inrush limit, fan supply, board voltagesNo power, tripping mains switch, missing rail shown on Internal I/O LEDs
07Vivid i Probe and MUX board128 channels in, HV mux to 64, CW processor, pencil probe pathProbe not recognised, dead elements on all probes, no CW
08Vivid i TR32 boards and RFI board64 channels with beamformers; front end control and processingFailed front end tests, noise or no image on every probe
09Vivid i BEP, hard disk and protection plugETX PC, software and data, option licencesNo boot, freezes, options lost after disk or BEP change
10Vivid i power supply, battery, adapter and fansSource selection, LV and HV rails, charging, coolingNo power-up, short run time, thermal shutdowns, blocked vents
Fault-finding flowIdentify the Vivid before you service it
A Vivid 7 or Vivid i arrivesfor service and you have notyet confirmed what it is.Do the rating labeland part number matchthe documents?1NoRecord the mismatch; clarifyownership and history before anywork.YesOn a Vivid 7, are thebuild level and BEPgeneration known?2NoFind them first; board revisions andsoftware discs depend on them.YesIs the softwareversion valid for thishardware?3NoDo not reload or upgrade until theright package is confirmed.YesDo the installedoptions match thecustomer's contract?4NoNote the gap; options are settledwith GE, not restored by you.YesAre probes,peripherals and cartrecorded with serials?5NoComplete the record; it is yourbaseline for every later check.YesWrite the baseline down:model, build, software,options, probes, peripherals,network.
Trainer: walk the fault-finding flowAnswer each check as if you were at the machine. The flow leads you to the cause.
field case

Worked example: the transmitter board that did not fit the build

This is a teaching example built from the Vivid 7 manual, not a field report. A Vivid 7 shows weak transmit on every probe, and the board tests point at the transmitter. The engineer orders a TX128 board by name. The manual's revision table, however, lists two transmitter versions: one for units produced before a given date in 2002 and for the Vivid 7 PRO, and a later version as the replacement for units produced after that date. It also lists different acceptable revisions of several other boards for the BT'01 and BT'02 build levels. Before ordering, the engineer reads the part number and build level of the console, checks the production date and the revision of the board that is fitted, and orders the version that the table allows for this unit.

On a Vivid 7 the board name is not enough. Match every replacement board to the build level, production date and revision table in the manual for that unit, and record what you fitted.

From service practice

Every soft key died at once: the PC battery came first

On a cart system of the later Vivid S5/S6 generation, all the active keys stopped responding together, while the system still started. Colleagues reminded the engineer that the back end of these scanners is a PC with a small backup battery for its clock and BIOS settings, and that a flat battery or lost BIOS settings can produce strange, system-wide symptoms. The advice was to replace that battery, enter the BIOS setup per the documentation and save the settings, and to check the hard disk, but not to run an automatic disk repair, because it can damage the patient database.

The back end of a Vivid is a PC, and PC faults look like scanner faults. When a symptom is system-wide and sudden, check the PC basics (backup battery, BIOS settings, disk) before blaming the keyboard or the front end, and protect the database while you do.

On-site checklist

  • Rating label, part number and voltage version recorded
  • Vivid 7: build level (BT'01 or BT'02), BEP generation and software version recorded
  • Vivid i: catalogue number, software version, SafeLock cart or stand-alone recorded
  • Board revisions compared with the manual's table before ordering parts
  • Probes with serial numbers, including TEE and pencil probes
  • Installed options compared with the customer's contract
  • Peripherals, network and DICOM destinations noted

Check yourself

1. A Vivid 7 gives a noisy image only on the probe in the left connector; the same probe works in the middle connector. Which assembly carries the connectors and selects between them?
2. An engineer used to the LOGIQ e expects a single master board in a Vivid i. What does the Vivid i front end actually consist of?
3. Before ordering a replacement beamformer board for a Vivid 7, what must you establish first?
4. A customer's Vivid i shows no CW Doppler with the pencil probe, while PW and imaging with the phased array are normal. Which path should you suspect first?
5. Why does a repair near the ECG input of either Vivid end with an ECG leakage test?
Next lesson →