Early access · courses are being reviewed by field engineers
Service EngineerAcademy
01Lesson 1 / 6 · 20 min

Imagers in the imaging department: three technologies, one film path

Recognise which kind of imager you are standing in front of, name its main assemblies in the order the film meets them, and say which assembly a complaint most likely belongs to.

Why a hospital still prints film

Most departments now read images on screens and store them in PACS, yet the film printer is still on the call list of almost every service engineer. Surgeons want a film on the light box in theatre, patients are discharged with films to take to another hospital, some clinics have no viewing stations at all, and in many places the law or the payer still expects a hard copy. Mammography, orthopaedics and dental clinics often print every day. When the printer stops, the scanners keep working, but nobody leaves the department with images, and the call arrives as an emergency.

A medical imager (the industry also says laser camera, dry camera, hard-copy camera or film printer) receives an image over the network, almost always as a DICOM print job, turns pixel values into optical density on a sheet of film, and delivers a dry sheet ready to read on a light box. The engineering problem is always the same: put a controlled amount of blackening at each point of the sheet, at the right place, and do it the same way every day.

There is a second, smaller family you will meet at the same time: video printers next to ultrasound scanners and endoscopy towers. They print on thermal paper or small thermal film from a video or USB signal. They use the same thermal-head principle as the larger direct thermal imagers, so much of this course applies to them too.

Three technologies you will find on site

Direct thermal dry imagers write the image with a thermal print head: a long row of tiny resistive heating elements across the width of the film. The film carries a heat-sensitive layer; where an element heats it, the layer darkens, and more heat gives more density. The image is formed and fixed in one pass, with no light and no separate development step. Agfa Drystar models and Sony UP-DF thermal film printers are examples from the field cases in this course, and the small video printers work on the same idea with paper.

Laser photothermographic dry imagers work in two steps. First a laser diode, scanned across the sheet by a rotating polygon mirror while the film moves under it, exposes a light-sensitive layer and leaves an invisible latent image. Then the sheet is wrapped around a heated drum or carried over a heated plate, and the heat develops the latent image into a visible one. No liquid chemistry is used and no silver solution leaves the machine. Carestream (Kodak) DryView, Konica Minolta DRYPRO and Fujifilm DRYPIX imagers are typical examples; the DRYPRO SIGMA service manual is the example machine we refer to in this course.

Legacy wet laser imagers expose conventional silver-halide film with a laser and then pass it through a wet processor with developer, fixer, wash and dryer, either built in or docked to the camera. You will still find a few in service, usually old and short of parts. They add chemical handling, silver recovery and processor maintenance to everything in this course. If you meet one, treat the processor as a separate machine with its own procedures and safety data sheets.

  • Direct thermal: heat writes the image directly; no light sensitivity in normal room light, but sensitive to heat and pressure.
  • Laser photothermographic: light writes a latent image, heat develops it; the unexposed film must be protected from light.
  • Wet laser: light writes, chemistry develops; add processor and chemical safety to your checklist.

Film: the consumable that causes most of the trouble

Each technology uses its own film, and the films are not interchangeable. A thermal film will not record a laser exposure, a laser film fed through a thermal printer produces nothing useful, and even within one brand there are different film types for general radiology and for mammography, on blue or clear base. The imager usually identifies the film from a tag on the box, the cartridge or the backing sheet: a barcode, an RF tag or a memory chip that tells it the film type, size, emulsion batch and sometimes how many sheets are in the pack. That tag is also how the imager chooses the right calibration data.

From this one fact come a whole set of field faults. Film loaded with the wrong side towards the head or laser prints nothing or only a faint silhouette. Film loaded without changing the backing sheet makes the imager think the pack is empty or overused. A box from another region, an unsupported type, or a non-genuine batch may be refused or misidentified. An expired or badly stored box gives high base density, which the imager often reports as a calibration failure. Before you suspect the head, the laser or a board, look at the box: type, side, expiry date, how it was stored and whether the tag or backing was replaced with the film.

Storage conditions are set by the film manufacturer and printed on the box or its data sheet: temperature, humidity, keeping away from heat sources, ionising radiation and chemical fumes, and for light-sensitive film protection from light until it is inside the imager. Check them on every visit. A film store next to a radiator or on a sunny windowsill causes faults that no repair will cure.

The main assemblies in the order the film meets them

Every dry imager, whatever the brand, follows the same path, so learn it as a chain. The film supply holds one or more trays or cartridges, each with a sensor that knows whether it is present and closed, and a tag reader. A pick-up mechanism, usually suction cups on an arm with a small vacuum pump, or a friction roller, lifts the top sheet and hands it to feed rollers. Transport rollers, belts and guides carry the sheet past sensors that confirm its leading and trailing edges at each stage.

In a thermal imager the sheet then passes between the thermal head and a rubber platen roller, line by line, while a drive motor and belt move it at a precisely controlled speed. In a laser imager the sheet is carried on an exposure transport under the optics module, where the laser beam sweeps across it, and then into the processor: a heated drum or plate held at a tightly controlled temperature, with rollers that keep the film in contact. After development the film passes a cooling section so that it does not curl or mark, and then a densitometer, which measures the density of test patches, before it drops into the output tray.

Around the path sit the electronics: a power supply, a machine control board for motors, heaters and sensors, an image processing computer with its storage (hard disk, flash card or similar) and network port, and a front panel or display. Many laser imagers also have an odour or charcoal filter and fans that remove the small amount of vapour released when the film is heat-developed. That filter is a PM item, and a forgotten one leaves deposits inside the machine.

Putting the complaint on the chain

When the department calls, your first job is to place the symptom on this chain before you open a cover. A message about the tray, the film count or the film type belongs to the supply and the tag. A jam with an area number belongs to transport. A line along the direction of travel in a thermal printer points at one or more head elements; a band across the film points at the drive and speed. Low or uneven density on a laser imager points at the processor temperature, the film or the calibration before it points at the laser. A printer that prints nothing at all while the modality says the job was sent belongs to the network and configuration side.

The same chain tells you who owns the fix. Film type, storage and loading belong to the department and the supplier, and the fix is training and purchasing. Transport, head, optics and processor belong to you or the authorised service. Network and DICOM settings belong partly to you and partly to whoever configures the modality and PACS. Write down which link you think is at fault and why: that note is the start of your service report.

Unit breakdownA dry imager along the film path
PartWhat it doesHow it fails
01Film supply (trays or cartridges)Holds film; detects presence, closure and sizeTray not detected, cover not closing, film fogged by an open tray
02Tag reader (barcode, RF tag, chip)Reads film type, size, batch and sheet countOld backing sheet left in, foreign or unsupported film refused
03Pick-up (suction cups, vacuum pump, arm)Lifts one sheet from the packHard or cracked cups, weak pump, two sheets picked, pick-up retries
04Feed and transport rollers, belts, guidesCarry the sheet along the path at set speedDust, glazed or worn rubber, weak springs, wrapped film, jams
05Path sensorsConfirm leading and trailing edges at each stageDust on optical sensors gives false jams and timing errors
06Thermal head and platen (thermal type)Heats the film line by line to make densityDead or worn elements give lines; debris baked on the head
07Laser optics and polygon (laser type)Scans the beam across the film to expose itLaser power falls with age; polygon motor errors; dust on optics
08Processor drum or plate (laser type)Develops the latent image by controlled heatWrong temperature, heater or sensor faults, film wrapped on drum
09DensitometerMeasures test-patch density for calibrationDirty optics or guide, cable faults give calibration errors
10Controller, storage, network portReceives DICOM jobs, renders pages, runs the machineWorn disks or flash cards, lost settings after power cuts
Fault-finding flowThe imager 'does not print properly': find the link first
The department reports thatfilms come out wrong or theimager will not print.Is the correct filmtype loaded, in date,right side up?1NoFilm problem: reload correctly orreplace the box; check how it wasstored.YesWas the backing sheetor tag replaced withthe new film?2NoSupply/tag mismatch: reload as afull pack per the operator manual.YesDoes the imager reachReady with no errorafter warm-up?3NoNote the code and its group; go tothe lesson for that subsystem.YesDoes the imager's owntest print lookcorrect?4NoFault is inside the imager:transport, head/optics, processor,densitometer.YesDoes a job from eachmodality arrive andprint?5NoNetwork or DICOM configurationproblem on the sending side (lesson5).YesCompare the clinical printwith the test print; if bothare good, record the visit andclose.
Trainer: walk the fault-finding flowAnswer each check as if you were at the machine. The flow leads you to the cause.
field case

The dry imager that was never switched off

A Konica Minolta DRYPRO SIGMA laser imager had been left running day and night for about three years, because earlier software crashes had made everyone afraid to power it down. Films had begun to show marks, and the engineer called for a routine visit found the odour filter heavily loaded with yellow sulphur-like deposits. The same deposit had built up in the air labyrinth inside the left cover and on the cooling area after the drum. The filter had not been changed for a long time. The engineer replaced the filter, opened and cleaned the labyrinth, and cleaned the cooling zone and the drum with isopropyl alcohol following the manufacturer's PM procedure. He then set up the filter change on the schedule the manufacturer gives in sheets printed, and recorded the drum and filter condition in the report.

The heat development of dry laser film releases vapour that the filter must catch. A filter treated as optional turns into deposits on the drum, the rollers and the film.

From service practice

Agfa Drystar 3000: blue sheets with only a silhouette

An Agfa Drystar 3000 direct thermal imager worked normally for half a day and then began to deliver sheets that came out blue with only a faint outline of the image. The first engineer on site assumed the thermal head had failed and replaced it, but the next sheets looked exactly the same. Someone also suggested a laser fault, which made no sense on a thermal printer. The real cause was found only when the film was examined: the pack had been put into the tray the wrong way, so the heat-sensitive side was facing away from the head. After the film was reloaded correctly, the prints were normal.

A sudden change after a film reload is a film-loading problem until proven otherwise. Check the side and the type of the film before you order a head.

From service practice

Agfa imager: a mystery extra sheet and the wrong film type

After each study an Agfa dry imager printed an extra sheet carrying a message asking the user to contact service, and it identified the DT5B film loaded in it as a different type, DT2B. Nothing else in the machine looked wrong. The engineer checked when the film had been bought and changed, and then asked the manufacturer to check the batch number from the barcode on the box. The batch was not genuine: the film had come through an unofficial supplier. The department returned it, bought genuine film, and after the change the imager identified the film correctly and the extra sheets stopped.

The imager trusts the film tag. Strange film-type messages after a new delivery point at the film and its source before they point at the electronics.

On-site checklist

  • Identify the technology: thermal head, laser plus heated processor, or wet processor.
  • Write down make, model, serial number, total print count and prints since last PM.
  • Check the film in use: type, size, expiry date, side, and whether the tag or backing was changed.
  • Look at where the film is stored and compare with the conditions on the box.
  • Ask what exactly is wrong: no print, jam, lines, bands, density, wrong size, wrong patient.
  • Place the complaint on the film path before you open any cover.

Check yourself

1. An imager writes the image with a row of tiny heating elements in a single pass, with no separate development stage. Which type is it?
2. After the night shift reloaded film, a direct thermal imager produces sheets with only a faint silhouette. The head was fine yesterday. What do you check first?
3. Why must the unexposed film of a laser photothermographic imager be protected from light, while thermal film is handled in room light?
4. On a laser dry imager, where does the visible image appear?
5. A laser imager smells stronger than usual and a yellow deposit is building up near the cooling area. Which PM item has most likely been missed?
Next lesson →